• 2.1 kW (9.6 A)
    Cat.No. Model Measurement wavelength (λ) Measurement range Temperature control range Speed control range
    R11-650-025 RTS-8 plus 850 ± 15 nm 0-100 OD600 +15 ℃ below ambient ... +60 ℃ 150-2700 rpm
    Specification
    Cat.No. R11-650-025
    Model RTS-8 plus
    Light source Laser
    Measurement wavelength (λ) 850 ± 15 nm
    Measurement range 0-100 OD600
    E.coli factory calibration measurement range 0-50 OD600
    S.cerevisiae factory calibration measurement range 0-75 OD600
    Achievable user calibration measurement error (range 0.1-6 OD600) ± 0.3
    Achievable user calibration measurement error (range 6-50 OD600) ≤ 5%
    Achievable user calibration measurement error (range 50-75 OD600) ≤ 10%
    Measurement periodicity per hour 1-60
    Temperature setting range +15℃ ... +60℃
    Temperature control range +15 ℃ below ambient ... +60 ℃
    Temperature stability ±0.3 ℃
    Sample temperature accuracy (20℃ -37℃) ±>1℃
    Tube sockets 8
    Sample working volume range 3-50 ml
    Speed control range 150-2700 rpm
    Reverse spin time setting range 150-250 rpm 0 s
    Reverse spin time setting range 250-300 rpm 2-60 s
    Reverse spin time setting range 300-2700 rpm 0-60 s
    Display LCD
    Minimum PC requirements Intel/AMD Processor, 1 GB RAM, Windows Vista/7/8/8.1/10, USB 2.0 port
    Dimensions (W×D×H) 350×690×300 mm
    Weight 20 kg
    Nominal operating voltage AC 230 V, 50 Hz
    Power consumption 3.15 A / 500 W
    ** **
    02 sensor +
    Range 0-100%
    Accuracy ±0.05% O2 at 0.2%,
    ±0.4% O2 at 20.9%
    Drift <0.03% O2 within 30 days
    Temperature range up to 40℃
    Response time (t90) <6 s
    Storage stability 18 months
    ** **
    pH sensor +
    Range 4.0 - 8.5 pH
    Accuracy ±0.10 pH at pH 7
    Drift <0.005 pH per day
    Temperature range up to 40℃
    Response time (t90) <120 s
    Storage stability 18 months