• Monochromator
    Cat. No. Model Spectral Bandwidth
    S23-3269-001 OPTIZEN POP 1.8 nm (190 to 1100 nm)
    S23-3269-021 OPTIZEN POP-V 3.0 nm (340 to 1100 nm)
    Cat. No. S23-3269-001 S23-3269-021
    Model OPTIZEN POP OPTIZEN POP-V
    Photometrics System Single-beam type
    Light Source Tungsten Halogen Lamp & Deuterium Lamp
    (Built-in light source auto interchanging motor)
    Tungsten Halogen Lamp
    Detector Silicon Photodiode
    Spectral Bandwidth <1.8 nm <3.0 nm
    Wavelength Range 190 to 1100 nm 340 to 1100 nm
    Wavelength Display(setting) 0.1 nm
    Wavelength Accuracy <¡¾ 0.5 nm
    Wavelength Repeatability <¡¾ 0.1 nm <¡¾ 0.2 nm
    Slew Rate About 7,800 nm/min
    Scan Speed max 4,000 nm/min
    Photometric Range Absorbance : -3 to 3 Abs
    Transmittance : 0 % to 300 %
    Photometric Accuracy <¡¾ 0.005 A (at 1.0 A), <¡¾0.003 A (at 0.5 A)
    Photometric Repeatability ¡¾ 0.001 A
    Stray Light < 0.1% T (220, 340 nm)
    Baseline Stability <¡¾ 0.001 A/h (at 700 nm) <¡¾ 0.002 A/h (at 700 nm)
    Baseline Flatness <¡¾ 0.001 A (200 -1100 nm) <¡¾ 0.003 A (340 to 1050 nm)
    Monochromator Czerny-Turner type with 1200 lines/nm blazed grating
    Lamp Interchange Wavelength 340~410 nm (Default 370 nm)
    Standard Cell Holder Automatic Rotary type 8-position Multi-Cell Holder
    Display 7 inch color LCD with touch screen
    Operating System(OS) Windows CE
    Power Requirement 100~240 V; 50~60 Hz
    Weight(kg) 8 kg
    Dimensions(W£øD£øH) 433 mm£ø381 mm£ø180 mm